SKZ161D E F Solar Radiation Simulation Tester

Product

SKZ161D E F Solar Radiation Simulation Tester

Product Download

If you want to download this product information Please click here

Used for testing the performance of discoloring, fading, aging, transmissibility, peeling, hardening,
softening, of the products and materials in the industries of rubber, plastic, paint, petrochemicals, auto,
fabric, for solar radiation simulation test( illumination, raining, temperature, humidity, )
Relevant Standards
ISO4892, ISO03917, ISO11341, ISO105, AATCC M16, AATCC169, ASTM G26, ASEJ1885KJ,
ASEJ1960KJ, GJB 150.7.

SKZ161D E F Solar Radiation Simulation Tester1

SKZ161D E F Solar Radiation Simulation Tester2

SKZ161D E F Solar Radiation Simulation Tester3

 

Instrument features:
1. an advanced computer controller and multi-channel temperature and humidity control system.
2. easily create complex test procedures to facilitate the choice and control light, temperature, humidity
and spray conditions.
3. menu-driven display guides the user through the button to select the test parameters.
4. button to select the database standard or user-defined test procedures.
5. using a short arc, high light intensity, water-cooled xenon arc lamp and professional double-filter system
simulate the real source of sunshine.
6. European imports of fixed-band spectrum traceable calibration systems, digital set light intensity,
real-time monitoring, automatically adjusted to meet different standards for light stability testing requirement.
7. superconducting fiber imported, exported by the whole band monitoring light source, light detector at
warehouse outside monitoring, no drift, no decay, high precision, life permanently.
8. hybrid water cooling system, greatly reducing water consumption and power consumption.
9. large experimental warehouse design, to maximize the exposure area.
10. the board thermometer (BPT), the standard blackboard Thermometer (BST) and sample the same
station (isometric) testing, a true reflection of the sample under test conditions, the measured CPU
processing digital data, charts, graphs, etc. displayed on the color screen, do not stop observing.
11. ultrasonic atomizer and PTC double humidification mode, professional dehumidification system to
ensure accurate and stable humidity test.
12. equipped with self-circulation system and air filtration system, significantly reducing the environmental
requirements.
13. 10.4-inch touch screen display control, monitoring a variety of test modes (animation, digital,
graphics), easy manipulation, clear and intuitive.
14. sample holder to achieve all the time, respectively, can be machine experiments with different samples,
to facilitate monitoring of the test.
15. front and back of the spray model.
16. the instrument is equipped with parallel communication interface, can provide Chinese A4 printing
(optional).
17. a test run of 1000 hours of continuous quality assurance.
Technical parameters 

 

<table “style=”height: 727px;” border=”1″ width=”671″ cellspacing=”0″ cellpadding=”0>

 Model
SKZ161D
 Test compartment temperature control
 30~70℃, Resolution: 0.1 ℃
 Storage humidity control test
 Ming period :10-75% RH, humidity fluctuations ± 3% RH; dark cycle :10-95% RH, humidity fluctuate ± 5% RH;
 Time control experiments
 0min~999:59h, Accuracy ± 1min
 Irradiance control
 1.00-1.30W/m2/420nm Accuracy: ± 0.02 W/m2/420nm;digital set,
automatic compensation.
 Control and monitoring wavelength
 Optional 340nm, 420nm ,300-400nm ,300-800nm
 Xenon arc lamp rated power
 4.5KW
 Sample holder rotation speed
 1rpm can be set
 Diameter distribution of sample clip
 φ500mm
 Maximum exposure area
 3500cm2
 Sample holder can be mounted with the
number of dimensions
 145 × 75mm model 25; 145 × 45mm model 35
Timing of each sample holder, respective
 ≤10000h
 Photoperiod
 ≤10000h
 Spray cycle
 ≤10000h
 Blackboard temperature range
 (BPT) 30-90℃±2℃;(BST) 35-95℃±2℃
 Display control
 10.4-inch touch screen
 Data Output
 Digital color display (A4 print in English optional)
 Sample holder type
 Single
 External circulating water requirements
 Water pressure :140-1345kPa          Flow: 1.0L/min
 Requirements within the recycled water
 High water flow: 0.5L/min
 Power supply
 AC380V±5% 50Hz 10KW
 Dimensions
 1250×830×1800mm
 Weight
 350kg
  Technical parameters
 Model
 SKZ161E
 Test chamber temperature control
 30~70℃, Distinguishability:0.1℃
 Test chamber humidity control
 light cycle : 10-75%RH, humidity fluctuation ±3%RH;
shade cycle :10-95%RH, humidity fluctuation ±5%RH;
 Testing time control
 0min~999:59h, Accuracy:±1min
 Irradiance
 0.8~2.5W/m2/420nm  Accuracy:±0.03W/m2/420nm, Digital setting
automatic compensation
 Control and monitoring wavelength
 Optional 340nm, 420nm, 300-400nm, 300-800nm wave band
 Rated power of xenon lamp
 4.5KW
 Slewing speed of sample frame
 1-10rpm settable
 Distribution diameter of sample clamp
 φ550mm
 Maximum exposure area
 3500cm2
 size and number of sample clamps
 145×75mm template 35 block;145×45mm template 48 pieces
 Each sample folder times respectively
 ≤10000h
 Illumination cycle
 ≤10000h
 Spraying cycle
 ≤10000h
 Blackboard temperature range
 (BPT) 30-90℃±2℃;(BST) 35-95℃±2℃
 manipulation showing
 10.4 inch touch screen
 Data output
 Digital color display (A4 print in English optional)
 Sample holder type
 Double
 External circulating water requirements
 Water pressure :140-1345kPa       Flow: 1.3L/min
 Requirements within the recycled water
 High water flow: 0.6L/min
 power source
 AC380V±5% 50Hz 10kW
 external dimensions
 1250×1000×1800mm
 Weight
 400kg
  Technical parameters 

<table “style=”height: 727px;” border=”1″ width=”100%” cellspacing=”0″ cellpadding=”0>

 Model
SKZ161F
 Test compartment temperature control
 20~93℃,  Resolution: 0.1 ℃
 Storage humidity control test
Light cycle:10-75% RH, humidity fluctuations ± 3% RH;
dark cycle :10-95% RH, humidity fluctuate ± 5% RH;
 Time control experiments
 0min~999:59h, Accuracy ± 1min
 Irradiance control
 0.8 ~ 2.5W/m2/420nm Accuracy: ± 0.03W/m2/420nm;digital set,
automatic compensation
 Control and monitoring wavelength
 Optional 340nm, 420nm ,300-400nm ,300-800nm
 Xenon arc lamp rated power
 6.5KW
 Sample holder rotation speed
 1-10rpm can be set
 Diameter distribution of sample clip
 φ650mm
 Maximum exposure area
 6900cm2
 Sample holder can be mounted with the
number of dimensions
 145 × 75mm model 68; 145 × 45mm model 90
 Timing of each sample holder, respective
 ≤10000h
 Photoperiod
 ≤10000h
 Spray cycle
 ≤10000h
 Blackboard temperature range
 (BPT) 30-100℃±2℃;(BST) 35-105℃±2℃
 Display control
 10.4-inch touch screen
 Data Output
 Digital color display (A4 print in English optional)
 Sample holder type
 Three
 External circulating water requirements
 Water pressure :140-1345kPa        Flow: 1.5L/min
 Requirements within the recycled water
 High water flow: 0.8L/min
 Power supply
 AC380V±5% 50Hz  12kW
 Dimensions
 1350×1050×1900mm
 Weight
 460kg

 

Name
*
Mobile
*
Email
*
Content
*