			Used to test the highest temperature called brittleness temperature when the specimen is broken by impact in prescribed conditions. Applied for non-rigid plastic and other flexible materials at low temperatures and also test the brittleness temperature and low temperature performance for different rubber material or different formula vulcanized rubber.
Standard:
GB1682-82, ISO812:1991, ISO974—80.
Technical parameters:
| Testing temperature | -80℃~0℃ | 
| Impact speed | 2 m/s±0.2 m/s | 
| Impactor weight | 200 ± 20g | 
| Working stroke | 40 ± 1mm | 
| Temperature fluctuation | <±0.5℃ | 
| Distance between impactor and clamp | 11±0.5 (mm) | 
| Distance between impactor and sample | 25±1mm | 
| Volume of colder | 3L | 
| Sample size | Length: 25.0±0.5mm, width: 6.0±0.5mm, thickness: 2.0±0.3mm | 
| Power | 1100W | 
| Dimension | 900*510*950mm | 
| Weight | 130KG | 
 
  
  
  
 